Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 2645A NetDAQ: CAL ADJ RS-232C /5700 INSTRUMENT: Fluke 2646A NetDAQ: CAL ADJ RS-232C /5700 DATE: 21-Oct-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 236 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # This procedure is based on the calibration procedure in the Imbedded # Software Requirements Specification dated March 22 1994, Rev 1.1. # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # ini_val.exe Gets default serial port for UUT from METCAL.INI. # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 DRAW 7 78 7 78 100 30 50 95 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Connect the UUT to an ac power source and turn it on. 1.005 DISP 1.005 DISP Warm-up time: 30 minutes. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 70%. 1.006 DOS ini_val metcal.ini startup port 1.007 DISP Connect the UUT to [MEM2]. 1.008 DISP To select the proper RS-232C parameters on NetDAQ press 1.008 DISP SHIFT and then COMM (LIST). Use the up and down arrows 1.008 DISP to make each selection and then press ENTER. 1.008 DISP 1.008 DISP [32] 4800 baud 1.008 DISP [32] no PARity 1.008 DISP [32] CTS On 1.008 DISP [32] Echo Off 1.009 HEAD PERFORMING SELF TEST... (~10s) 1.010 PORT [P4800,N,8,1,H][T20000] 1.011 PORT [CLR]*TST?[10][D5000][I] 1.012 MATH MEM1 = MEM 1.013 JMPZ 1.016 1.014 HEAD SELF TEST FAILED: STATUS = [MEM] 1.015 JMP 1.160 1.016 HEAD EQUIPMENT SETUP 1.017 DISP 1.017 DISP *************************************************** 1.017 DISP 1.017 DISP To reduce noise pickup by test leads, particularly 1.017 DISP during high Ohms verification, use shielded test 1.017 DISP cables between the Fluke 5700A and Fluke NetDAQ. 1.017 DISP 1.017 DISP *************************************************** 1.018 DISP Make the following connections to the NetDAQ: 1.018 DISP [32] 5700A OUTPUT V Ohm A HI to Input Module CH1 HI 1.018 DISP [32] 5700A OUTPUT V Ohm A LO to Input Module CH1 LO 1.018 DISP [32] 5700A SENSE V Ohm HI to Input Module CH11 HI 1.018 DISP [32] 5700A SENSE V Ohm LO to Input Module CH11 LO 1.019 DISP Make the following connections on the Input Module: 1.019 DISP [32] Input Module CH2 HI to Input Module CH12 HI 1.019 DISP [32] Input Module CH2 LO to Input Module CH12 LO 1.019 DISP [32] Input Module CH2 HI to Input Module CH2 LO 1.020 DISP The Cal Enable button is located on the right side 1.020 DISP of the display and is recessed beneath the calibration 1.020 DISP seal. [DRAW 50,90,F] 1.020 DISP Enable the calibration mode by pressing the 1.020 DISP Cal Enable button for about three (3) seconds. 1.020 DISP Use a small screwdriver or equivalent blunt-tipped 1.020 DISP object. Avoid sharper-tipped objects (such as pencils. 1.020 DISP ("CAL" appears in the display.) 1.021 DISP *************************************************** 1.021 DISP In case a time-out error occurs, verify that 1.021 DISP the Fluke NetDAQ is in the CAL mode! 1.021 DISP *************************************************** 1.022 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEPS 1-6 1.023 PORT [CLR]CAL 1;*OPC?[10][I] 1.024 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 1 1.025 5700 90.000mV S 2W 1.026 PORT [D7000][CLR]CAL_STEP?[10][I] 1.027 MATH MEM1 = MEM - 0.09 1.028 JMPZ 1.030 1e-006U 1.029 JMP 1.060 1.030 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 2 1.031 5700 0.3000V S 2W 1.032 PORT [D7000][CLR]CAL_STEP?[10][I] 1.033 MATH MEM1 = MEM - 0.3 1.034 JMPZ 1.036 1e-006U 1.035 JMP 1.060 1.036 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 3 1.037 5700 0.7500V S 2W 1.038 PORT [D7000][CLR]CAL_STEP?[10][I] 1.039 MATH MEM1 = MEM - 0.75 1.040 JMPZ 1.042 1e-006U 1.041 JMP 1.060 1.042 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 4 1.043 5700 3.0000V S 2W 1.044 PORT [D7000][CLR]CAL_STEP?[10][I] 1.045 MATH MEM1 = MEM - 3 1.046 JMPZ 1.048 1e-006U 1.047 JMP 1.060 1.048 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 5 1.049 5700 30.000V S 2W 1.050 PORT [D7000][CLR]CAL_STEP?[10][I] 1.051 MATH MEM1 = MEM - 30 1.052 JMPZ 1.054 1e-006U 1.053 JMP 1.060 1.054 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 6 1.055 5700 50.000V S 2W 1.056 PORT [D7000][CLR]CAL_STEP?[10][I] 1.057 MATH MEM1 = MEM - 50 1.058 JMPZ 1.064 1e-006U 1.059 JMP 1.060 1.060 HEAD [MEM]VDC CALIBRATION ADJUSTMENT FAILED! 1.061 OPBR Do you want to repeat DC Volts calibration adjustment? 1.062 JMPT 1.023 1.063 JMP 1.160 1.064 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEPS 7-12 1.065 5700 * S 1.066 PORT [CLR]CAL 2;*OPC?[10][I] 1.067 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 7 1.068 5700 30.000mV 1kH S 2W 1.069 PORT [D20000][CLR]CAL_STEP?[10][I] 1.070 MATH MEM1 = MEM - 0.03 1.071 JMPZ 1.073 1e-006U 1.072 JMP 1.102 1.073 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 8 1.074 5700 300.00mV 1kH S 2W 1.075 PORT [D20000][CLR]CAL_STEP?[10][I] 1.076 MATH MEM1 = MEM - 0.3 1.077 JMPZ 1.079 1e-006U 1.078 JMP 1.102 1.079 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 9 1.080 5700 300.00mV 1kH S 2W 1.081 PORT [D20000][CLR]CAL_STEP?[10][I] 1.082 MATH MEM1 = MEM - 0.3 1.083 JMPZ 1.085 1e-006U 1.084 JMP 1.102 1.085 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 10 1.086 5700 3.0000V 1kH S 2W 1.087 PORT [D20000][CLR]CAL_STEP?[10][I] 1.088 MATH MEM1 = MEM - 3 1.089 JMPZ 1.091 1e-006U 1.090 JMP 1.102 1.091 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 11 1.092 5700 3.0000V 1kH S 2W 1.093 PORT [D20000][CLR]CAL_STEP?[10][I] 1.094 MATH MEM1 = MEM - 3 1.095 JMPZ 1.097 1e-006U 1.096 JMP 1.102 1.097 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 12 1.098 5700 30.000V 1kH S 2W 1.099 PORT [D20000][CLR]CAL_STEP?[10][I] 1.100 MATH MEM1 = MEM - 30 1.101 JMPZ 1.106 1e-006U 1.102 HEAD [MEM]VAC CALIBRATION ADJUSTMENT FAILED! 1.103 OPBR Do you want to repeat AC Volts calibration adjustment? 1.104 JMPT 1.064 1.105 JMP 1.160 1.106 HEAD OHMS CALIBRATION ADJUSTMENT STEPS 12-16 1.107 5700 * S 1.108 PORT [CLR]CAL 3;*OPC?[10][I] 1.109 HEAD OHMS CALIBRATION ADJUSTMENT STEP 12 1.110 5700 190Z S 4W 1.111 PORT [CLR]CAL_REF [MEM1];*OPC?[10][I] 1.112 PORT [D15000][CLR]CAL_STEP?[10][I] 1.113 MATH MEM1 = MEM - MEM1 1.114 JMPZ 1.116 0.01U 1.115 JMP 1.144 1.116 HEAD OHMS CALIBRATION ADJUSTMENT STEP 13 1.117 5700 1.9e3Z S 4W 1.118 PORT [CLR]CAL_REF [MEM1];*OPC?[10][I] 1.119 PORT [D15000][CLR]CAL_STEP?[10][I] 1.120 MATH MEM1 = MEM - MEM1 1.121 JMPZ 1.123 0.1U 1.122 JMP 1.144 1.123 HEAD OHMS CALIBRATION ADJUSTMENT STEP 14 1.124 5700 1.9e4Z S 4W 1.125 PORT [CLR]CAL_REF [MEM1];*OPC?[10][I] 1.126 PORT [D15000][CLR]CAL_STEP?[10][I] 1.127 MATH MEM1 = MEM - MEM1 1.128 JMPZ 1.130 1U 1.129 JMP 1.144 1.130 HEAD OHMS CALIBRATION ADJUSTMENT STEP 15 1.131 5700 1.9e5Z S 4W 1.132 PORT [CLR]CAL_REF [MEM1];*OPC?[10][I] 1.133 PORT [D15000][CLR]CAL_STEP?[10][I] 1.134 MATH MEM1 = MEM - MEM1 1.135 JMPZ 1.137 10U 1.136 JMP 1.144 1.137 HEAD OHMS CALIBRATION ADJUSTMENT STEP 16 1.138 5700 1.9e6Z S 4W 1.139 PORT [CLR]CAL_REF [MEM1];*OPC?[10][I] 1.140 PORT [D15000][CLR]CAL_STEP?[10][I] 1.141 MATH MEM1 = MEM - MEM1 1.142 JMPZ 1.148 100U 1.143 JMP 1.144 1.144 HEAD [MEM]j CALIBRATION ADJUSTMENT FAILED! 1.145 OPBR Do you want to repeat Ohms calibration adjustment? 1.146 JMPT 1.108 1.147 JMP 1.160 1.148 HEAD FREQUENCY CALIBRATION ADJUSTMENT STEP 17 1.149 5700 * S 1.150 PORT [CLR]CAL 4;*OPC?[10][I] 1.151 5700 1.0e4H 3.0V S 2W 1.152 PORT [D5000][CLR]CAL_STEP?[10][I] 1.153 MATH MEM1 = MEM - 10000 1.154 JMPZ 1.159 1e-006U 1.155 HEAD [MEM]Hz CALIBRATION ADJUSTMENT FAILED! 1.156 OPBR Do you want to repeat frequency calibration adjustment? 1.157 JMPT 1.150 1.158 JMP 1.160 1.159 HEAD CALIBRATION ADJUSTMENT COMPLETE 1.160 5700 * S 1.161 PORT [CLR]*RST;*OPC?[10][I] 1.162 END